QM 001 QualityManual
1. Purpose¶
Describe why ACME GmbH maintains a Quality Management System (QMS), what the QMS covers, how the controlled document set fits together, and how quality objectives are translated into operating procedures and records.
2. Scope¶
Applies to the ACME GmbH QMS for healthcare software products, including products that may qualify as software as a medical device (SaMD), and to the company-level governance baseline maintained in this upstream baseline repository.
3. Organization and Context¶
- Legal Entity: ACME GmbH
- Registered Address: Paradeplatz 8, 8001 Zurich, Switzerland
- Operating Model: Remote-first company
- Regulatory Role: Manufacturer
- Product Type: Healthcare software products; some products may qualify as SaMD
- Intended Use Policy: Each product must define intended use, indications, claims, and regulatory classification before development and release under controlled procedures.
- Infrastructure Policy: Each product must define and approve its infrastructure architecture and controls against applicable security, availability, and traceability requirements within the product dossier.
4. QMS Intent and Quality Policy¶
ACME maintains this QMS to:
- apply ISO 9001:2015+Amd 1:2024 quality management expectations for organization context, customer focus, process performance, and continual improvement
- ensure consistent compliance with ISO 13485:2016 and applicable regulatory requirements
- apply IEC 62304:2006+A1:2015 lifecycle and maintenance controls to regulated software
- apply ISO 14971:2019 risk management across the software lifecycle, including cybersecurity-related harms
- apply IEC 62366-1:2015 usability engineering for safety-related user interfaces
- apply IEC 82304-1:2016 product-safety controls for health software placed on general computing platforms
- apply ISO/IEC 27001:2022 information security governance for QMS, product, supplier, and operational information assets
- apply ISO/IEC 42001:2023 AI management controls for AI components inside ACME software products and AI-bearing supplier interfaces
- apply EN 18286:2026 provider QMS controls, through SOP-024, when ACME places a high-risk AI system on the EU market or puts it into service
- maintain product safety and performance through controlled lifecycle processes
- retain objective evidence and traceability for quality decisions
Quality Policy:
ACME GmbH commits to delivering safe and effective healthcare software through a risk-based, evidence-driven quality management system and continuous improvement.
The approved policy baseline is maintained in matrices/company_profile.yml.
The QMS uses a risk-based approach in which processes are defined, measured, and continuously improved.
5. Quality Objectives¶
The quality objective framework is maintained in matrices/company_profile.yml and is reviewed through management review. The current company-level objectives are to:
- merge controlled pull requests only after required approval on the current head commit
- release only baselines with complete V&V, risk, and release traceability evidence
- close critical quality actions on time
- complete required controlled-document training on time
- maintain management review cadence
- monitor customer feedback, satisfaction signals, and process-performance trends
- ensure products perform against defined performance criteria and complaint thresholds
- maintain information security and AI management controls for applicable QMS and product lifecycle scope
Each objective must have an accountable role, a measurable indicator, and a target threshold.
6. QMS Structure and Process Interaction¶
The QMS is structured as:
1. QM-001 Quality Manual: purpose, scope, policy, objectives, and system architecture.
2. SOP-005 QMS Governance: governance cadence, ownership, document hierarchy, communication, and governance interfaces.
3. Operational SOPs: controlled procedures for document control, CAPA, audit, management review, training, design control, change control, suppliers, feedback, incident, PMS, nonconformity, metrics, infrastructure, risk, usability, and software lifecycle control.
4. WIs: detailed execution instructions for selected lifecycle activities.
5. Matrices and records: traceability baselines, role/training assignments, and objective evidence.
High-level process interaction:
| QMS Area | Primary Controlled Documents | Main Outputs |
|---|---|---|
| QMS framing and policy | QM-001, SOP-005, SOP-004, SOP-016 |
Scope statement, policy, objectives, management review decisions |
| Document and training control | SOP-001, SOP-011 |
Controlled revisions, training assignments, training logs |
| Product lifecycle control | SOP-007, SOP-008, SOP-009, SOP-020, SOP-025, WI-001, WI-002 |
Product dossier, development plans, release baselines, V&V evidence, clinical evaluation and PMCF records |
| Risk and usability | SOP-018, SOP-019 |
Risk file, usability file, residual risk decisions |
| Suppliers and post-market processes | SOP-010, SOP-012, SOP-013, SOP-014, SOP-015 |
Supplier approvals, complaint/PMS/incidents, nonconformity handling |
| Improvement and oversight | SOP-002, SOP-003, SOP-004, SOP-016 |
CAPAs, audits, reviews, metrics and improvement actions |
7. Roles and Responsibility Overview¶
| Role | System-Level Responsibility |
|---|---|
| Management Representative | Acts for top management within the small-team model, approves the quality policy/objective baseline, and confirms QMS adequacy and effectiveness. |
| QA Lead | Maintains the controlled QMS baseline, owns regulatory pathway and dossier completeness expectations, verifies process interaction completeness, and drives quality-system improvements. |
| Engineering Lead | Owns lifecycle planning, technical release readiness, and engineering process implementation. |
| Usability Lead | Owns safety-related usability engineering inputs and file completeness for user-facing products. |
| Technical QMS Maintainer | Maintains quality tooling, automation, and technical QMS infrastructure under controlled change. |
| Auditor | Verifies conformance and effectiveness of selected QMS processes and must remain independent of the activities being audited. |
Detailed governance responsibilities are defined in sops/SOP-005-QMSGovernance.md.
8. Standards and Regulatory Framework¶
Regulatory applicability is maintained in matrices/regulatory_market_scope.yml. The QMS baseline is aligned to:
- ISO 9001:2015, including ISO 9001:2015/Amd 1:2024
- ISO 13485:2016 / EVS-EN ISO 13485:2016+A11:2021
- IEC 62304:2006+A1:2015
- ISO 14971:2019 / EVS-EN ISO 14971:2019+A11:2021
- IEC 62366-1:2015+A1:2020
- IEC 82304-1:2016
- ISO/IEC 27001:2022, including ISO/IEC 27001:2022/Amd 1:2024
- ISO/IEC 42001:2023
- EN 18286:2026 (EU AI Act provider QMS, implemented through SOP-024), for high-risk AI systems placed on the EU market or put into service
Jurisdiction references currently maintained in the regulatory scope baseline include:
- Switzerland: MedDO, including software classification references and the clinical evaluation requirement of MedDO Art. 21(4) and Art. 46(3) (implemented through SOP-025)
- European Union: MDR and MDCG software guidance, including MDR Art. 61 and Annex XIV clinical evaluation and PMCF (implemented through SOP-025), and Regulation (EU) 2024/1689 (AI Act) for high-risk AI systems
- United States: FD&C Act device definitions, section 520(o), FDA software guidance, and 21 CFR Part 820
9. Document Architecture and Record Model¶
Document-control rules are defined in SOP-001, governance rules in SOP-005, and role-specific training expectations in SOP-011 plus matrices/training_matrix.yml.
The record model is: - company-level QMS documents and selected quality records are maintained in a controlled central repository that constitutes the authoritative QMS baseline - standardized templates and controlled document structures are published from this repository to support consistent implementation across products and studies - product- and study-specific records are maintained in designated controlled repositories and linked to the relevant medical device file and release evidence
QMS tooling and validation baselines are maintained in:
- matrices/qms_tooling_inventory.yml
- records/validation/qms-tools/github-validation-record.md
10. Traceability¶
| Standard Clause | Quality Manual Coverage | Linked Baseline |
|---|---|---|
| ISO 13485:2016 4.1, 4.2.1, 4.2.2 | Defines QMS scope, structure, and interaction model | matrices/quality_manual_traceability.yml |
| ISO 13485:2016 5.1, 5.3 | Defines management commitment and quality policy context | matrices/company_profile.yml, sops/SOP-004-ManagementReview.md |
| ISO 13485:2016 5.4.1, 5.4.2 | Defines quality objectives and planning context | matrices/company_profile.yml, sops/SOP-016-QualityMetricsAndDataAnalysis.md |
| ISO 13485:2016 5.5 | Defines system-level responsibilities and communication overview | sops/SOP-005-QMSGovernance.md, sops/SOP-011-CompetenceTrainingAndAwareness.md |
| ISO 9001:2015 4-10 | Defines the general quality-management overlay for context, interested parties, customer focus, organizational knowledge, process performance, and improvement | matrices/iso_9001_gap_analysis.yml, sops/SOP-005-QMSGovernance.md, sops/SOP-012-FeedbackAndComplaintHandling.md, sops/SOP-016-QualityMetricsAndDataAnalysis.md |
| Regulation (EU) 2017/745 Art. 61, Annex XIV; ISO 13485:2016 7.3.7 | Defines clinical evaluation and PMCF for medical device products as part of design validation and post-market surveillance | matrices/mdr_clinical_evaluation_gap_analysis.yml, sops/SOP-025-ClinicalEvaluation.md |
| EN 18286:2026 4-10; Regulation (EU) 2024/1689 Art. 17 | Defines the AI Act provider QMS layer for high-risk AI systems placed on the EU market, applied when the EN 18286 commitment in matrices/regulatory_market_scope.yml is applied |
matrices/en_18286_gap_analysis.yml, sops/SOP-024-AIActProviderQMS.md |
11. Related Controlled Documents¶
- SOP-001 Document and Record Control
- SOP-004 Management Review
- SOP-005 QMS Governance
- SOP-011 Competence, Training, and Awareness
- SOP-016 Quality Metrics and Data Analysis
matrices/company_profile.ymlmatrices/quality_manual_traceability.ymlmatrices/regulatory_market_scope.yml
12. Revision History¶
| Revision | Revision Date | Change Summary |
|---|---|---|
| R00 | 2026-03-08 | Initial release split from SOP-005 so the QMS intent, scope, policy, objectives, and process structure are available as a standalone Quality Manual. |
| R01 | 2026-03-10 | Updated public upstream baseline naming, licensing, and repository references for open-source publication. |
| R02 | 2026-03-25 | Tightened company-level scope wording, added risk-based quality framing and product performance objectives, and clarified auditor independence plus repository-of-record language. |
| R03 | 2026-03-27 | Consolidated the standalone regulatory-lead baseline role into the QA-lead governance model and aligned the system-level role overview accordingly. |
| R04 | 2026-04-23 | Added IEC 82304-1:2016 health software product-safety scope to the QMS intent and standards framework. |
| R05 | 2026-04-27 | Added ISO/IEC 27001:2022 information security governance to the QMS intent and standards framework. |
| R06 | 2026-04-27 | Added ISO/IEC 42001:2023 AI management controls for AI components inside ACME medical-device software to the QMS intent and standards framework. |
| R07 | 2026-04-27 | Renamed YAML metadata field effective_date to revision_date to align controlled-document metadata with the tag-based publication model. The effective date is derived from the published QMS release tag rather than from the document front matter. |
| R08 | 2026-05-04 | Standardize standards citations to specific versions. |
| R09 | 2026-05-26 | Added the ISO 9001 general quality-management extension for organization context, customer focus, process performance, and continual improvement. |
| R10 | 2026-10-26 | Added EN 18286:2026 EU AI Act provider QMS and the EU AI Act to the standards and regulatory framework, with a traceability row cross-referencing SOP-024. |
| R11 | 2026-10-28 | Aligned the QMS intent with SOP-022 R03, so ISO/IEC 42001 AI management controls cover AI components in all ACME software products, not only medical-device software, and added the EN 18286:2026 provider QMS intent implemented through SOP-024. |
| R12 | 2026-11-19 | Added MDR Art. 61 and Annex XIV clinical evaluation and PMCF to the regulatory framework and process structure, with a traceability row cross-referencing SOP-025, and the corresponding MedDO clinical evaluation provisions for Switzerland. |