1. Purpose

Define how ACME identifies, evaluates, and reports adverse events/incidents to regulatory authorities within applicable jurisdictional timelines.

2. Scope

Applies to each product that qualifies as medical device software under applicable regulatory requirements. Regulatory applicability, including jurisdictions and frameworks (e.g. EU MDR, UK MDR/UKCA, Swiss MedDO, US FDA regulations, and MDSAP-participating countries such as Canada and Australia), is defined and maintained in the regulatory scope baseline. A reportable incident includes any event that has led, or could have led, to serious deterioration of health, serious injury, or death, or represents a significant malfunction that could recur.

3. Inputs

  • Complaint and feedback cases
  • PMS signal assessments
  • Product risk files and known hazard controls

4. Outputs

  • Reportability decision record
  • Regulatory submission package and submission evidence
  • Follow-up/corrective action linkage

5. Roles and Responsibilities

Role Responsibilities
Management Representative Approves reportability decisions and submissions.
QA Lead Detects and triages potential incidents and ensures process compliance and record completeness.
Engineering/Medical Experts Provide technical and clinical impact assessments.

6. Procedure

6.1 Event Detection and Immediate Escalation

  1. Flag any event suggesting serious deterioration, injury risk, or significant malfunction recurrence.
  2. Escalate potential reportable events to management representative without delay.

6.2 Reportability Assessment

  1. Assess event facts against jurisdictional reportability rules maintained in product dossier.
  2. Document decision rationale, including uncertainty and assumptions.
  3. For ambiguous cases, classify as reportable unless justified otherwise by qualified review.

6.3 Submission Preparation

  1. Compile required facts: event narrative, device/software identifiers, risk assessment, corrective actions, and known occurrences.
  2. Ensure consistency with complaint/CAPA/PMS records.
  3. Submission timelines shall be determined and met in accordance with applicable regulatory requirements.

6.4 Follow-Up and Field Actions

  1. Track follow-up requests from authorities.
  2. Link incident outcomes to CAPA and product change controls.
  3. Where needed, coordinate customer communication and field corrective action.

6.5 Trend Integration

  1. Feed incident data into PMS trend analysis and management review package.
  2. Reassess risk controls when incident pattern changes.

7. Required Records

  • Incident intake and reportability assessment
  • Regulatory submission evidence and authority correspondence
  • Linked CAPA/change and closure summary

8. Traceability

Standard Clause Control in this SOP
ISO 13485:2016 8.2.3 Defines regulatory reporting controls for applicable adverse events/advisory notices.
  • SOP-012 Feedback and Complaint Handling
  • SOP-014 Post-Market Surveillance
  • SOP-002 CAPA

10. Revision History

Revision Revision Date Change Summary
R00 2026-03-02 Initial full release.
R01 2026-03-07 Renamed the incident-process business owner role to the controlled product/regulatory owner role used in small-team deployments.
R02 2026-03-07 Simplified the incident-process business role name to regulatory_lead.
R03 2026-03-18 Removed top-table standards clause metadata; normative references remain in the Traceability section.
R04 2026-03-25 Expanded regulatory-scope applicability wording, defined reportable incidents, and clarified that submission timelines follow applicable regulations.
R05 2026-03-27 Consolidated the standalone regulatory-lead incident-owner role into the QA-lead baseline and aligned the role table plus ownership metadata.
R06 2026-04-27 Renamed YAML metadata field effective_date to revision_date to align controlled-document metadata with the tag-based publication model. The effective date is derived from the published QMS release tag rather than from the document front matter.
R07 2026-05-04 Standardize standards citations to specific versions.