跳转至

Medical Device File Index Template

This template shows one workable Markdown-first index for the product medical device file (MDF).

It can serve as the GitHub-native index for the product dossier / technical documentation set and, where a team uses the term, the design history file style artifact log.

You may rewrite the sections or tables to match the product repository, as long as the current controlled artifact set, revision state, and ownership remain clear.

Replace the example text below with product-specific content before approval.

1. Purpose and Scope

The medical device file index is the product-level index to the current technical documentation set. It identifies, for one product, the controlled artifacts that together form the medical device file required by ISO 13485 4.2.3 and the technical documentation expected under applicable medical device regulations (for example EU MDR Annex II/III).

The index is not the source of any individual artifact. Each row points to the current controlled record that lives elsewhere in the product repository. The index is updated whenever the underlying record set changes so that a reader, auditor, or release reviewer can locate the current state of the dossier from a single entry point.

Example:

  • Product: ACME-GENERIC-SAMD
  • Repository: github.com/acme/acme-generic-samd
  • Primary market scope: EU MDR first release, with Switzerland and US assessment tracked in linked regulatory decisions
  • Notes: the controlled artifact references that locate the current intended use, device description, classification, and traceability summary records are listed in Section 2 rather than duplicated in the YAML header

2. Current Controlled Artifact Index

Record one row per deliverable family or concrete controlled record set that belongs in the current product dossier.

# Dossier Area Current Record Path Revision / Baseline Status Gate / Release Primary Owner Notes
1 Intended use and claims records/mdf/intended_use.md R03 approved Gate 1 baseline qa_lead Controlled intended use, users, use environment, and claims
2 Device description records/mdf/device_description.md R02 approved Gate 1 baseline engineering_lead + qa_lead Product overview, major subsystems, interfaces, variants, and accessories
3 Classification and pathway records/mdf/classification_decision.md R01 approved Gate 1 baseline qa_lead Distinct from IEC 62304 software safety classification
4 Design and development plan records/design/design_development_plan.md R01 approved Gate 1 baseline engineering_lead Lifecycle model, gates, and linked plans
5 User needs records/design/user_needs.md R02 approved Gate 1 baseline product_manager Needs, use context, and critical user outcomes
6 System and software requirements records/design/system_requirements_spec.md R04 approved Gate 1 baseline engineering_lead Testable functional, safety, security, and usability requirements
7 Software architecture records/design/software_architecture_spec.md R03 approved Gate 2 engineering_lead Components, interfaces, external dependencies, and deployment model
8 Design traceability matrix records/design/design_traceability_matrix.md R02 approved Gate 2 / Gate 3 engineering_lead Links needs, requirements, risks, tests, and released baselines
9 Risk management plan records/risk/risk_management_plan.md R01 approved Gate 1 baseline qa_lead Defines methods, scoring, review cadence, and responsibilities
10 Risk file and reviews records/risk/risk_register.md, records/risk/risk_management_review.md R05 / R02 current Gate 2 / Gate 3 engineering_lead + qa_lead Current hazard, failure-mode, and residual-risk state
11 Verification and validation records/verification_validation/vv_plan.md, records/verification_validation/vv_report.md R02 / R01 approved Gate 2 / Gate 3 qa_lead + engineering_lead Formal campaign planning, execution, and release recommendation
11a Clinical evaluation records/clinical/clinical_evaluation_plan.md, records/clinical/clinical_evaluation_report.md, records/clinical/pmcf_plan.md, records/clinical/pmcf_evaluation_report.md R01 / R01 / R00 / N/A approved Gate 3 / post-release qa_lead + management_representative CEP, CER with supporting records (SOP-025), PMCF plan and reports; MDR Annex II 6.1(c), (d)
12 Release records records/configuration/release_plan.md, records/configuration/release_baseline_manifest.md, records/configuration/final_release_decision.md R02 / R01 / R01 approved Gate 2 / Gate 3 engineering_lead + qa_lead Candidate scope, released baseline, and final authorization
13 Declaration of conformity records/mdf/declaration_of_conformity.md R00 draft until placed on market release package qa_lead Signed only for the placed-on-market configuration and jurisdictions in scope
14 Accompanying information (IFU and technical description) records/mdf/accompanying_information.md R01 approved release package qa_lead + engineering_lead Controlled IFU, user-facing labeling, and technical description delivered with the released product version
15 Traceability summary records/mdf/traceability_summary.md R01 approved Gate 3 qa_lead High-level dossier completeness and linkage summary
16 Post-market and quality follow-up links records/feedback/, records/pms/, records/incidents/, records/capas/ current ongoing post-release qa_lead Linked lifecycle records, not Gate 1 baseline artifacts
17 Retirement / decommissioning record records/configuration/retirement_decision.md N/A not applicable until product retirement post-release qa_lead + engineering_lead Controlled retirement decision, user/customer impact, communication, residual-risk handling, and record retention per SOP-020

3. Release and Change History

Use this table to keep the dossier index aligned with released baselines and major controlled revisions.

Release / Change Date PR / Issue Main Records Updated Notes
v1.0.0 YYYY-MM-DD #123 design plan, requirements, architecture, risk file, V&V report, release records, DoC Initial regulated release
v1.0.1 YYYY-MM-DD #145 risk review, regression evidence, release baseline, release decision Post-release corrective patch

4. Periodic Completeness Review

Example:

  • Last completeness review date: YYYY-MM-DD
  • Reviewer roles: management_representative, engineering_lead, qa_lead
  • Missing or stale records: none | list here
  • Records requiring update before next release: none | list here
  • Follow-up issue or CAPA reference: #123 or N/A

5. Linked Post-Market and Quality Follow-Up

Example:

  • Complaint log reference: records/feedback/complaint_case_log.md
  • PMS plan/reference: records/pms/pms_plan.md
  • Incident assessment reference: records/incidents/reportability_assessment.md
  • CAPA / nonconformity references: records/capas/CAPA-001.md, records/nonconformity/NC-001.md

6. Approval Record

Example:

  • Index approved for current release baseline: yes
  • Approval date: YYYY-MM-DD
  • Linked approving PR: #123
  • Notes: approval confirms the index points to the current controlled artifact set for the released configuration