Clinical Evaluation Report Template¶
Implements MDR Art. 61(12), Annex XIV Part A section 4, and SOP-025 §6.10 to
§6.11. The structure covers the topics of the notified body clinical
evaluation assessment report (MDCG 2020-13), so a reviewer can find and check
each topic. Write it for a clinical reader. Summarise and appraise sources
rather than copying them, and reference every source so it can be sampled.
1. Administrative Particulars¶
- Manufacturer, device name, version evaluated, UDI-DI, class, conformity assessment route:
- CER version, date, and the CER it supersedes:
- Evaluators, with links to declarations:
- Summary of changes since the last CER:
2. Executive Summary¶
State the conclusion on safety, performance, clinical benefit, and benefit-risk in a few paragraphs, listing the claims supported and any claims removed.
3. Scope and Device Description¶
- Intended purpose, indications, contraindications, population, users, environment (consistent with
records/mdf/intended_use.md): - Device description from a clinical-use point of view: inputs, outputs, and role in the clinical pathway (include a workflow figure):
- Variants, configurations, accessories, platforms:
- Novel features:
- Expected lifetime and factors that may change performance:
- Classification and rationale:
4. Clinical Evaluation Plan¶
- Reference to the approved CEP and any revisions, with reasons:
- GSPRs requiring clinical data:
- Claims, endpoints, and acceptance criteria (table from the CEP):
5. State of the Art¶
Summary of the state of the art assessment, the benchmarks derived, and their origin.
6. Data Identification¶
- Literature searches: protocol reference, dates, results flow, inclusion and exclusion outcome:
- Manufacturer-held data: analytical, verification and validation, usability, simulated-use, clinical studies:
- PMS, PMCF, vigilance, and complaint data:
- Equivalence (if used): conclusion and reference to the equivalence assessment:
- Art. 61(10) (if used): conclusion and reference to the justification:
7. Appraisal¶
Summary of the appraisal method and outcome per data set, including the data
type (clinical data versus technical or analytical data) and weight. Reference
records/clinical/clinical_data_appraisal.yml.
8. Analysis¶
8.1 Evidence chain¶
| Claim / benefit | Valid clinical association | Technical performance | Clinical performance | Clinical benefit | Conclusion |
|---|---|---|---|---|---|
8.2 Performance¶
Results against each acceptance criterion and the state-of-the-art benchmark. For AI-enabled software, include subgroup and edge-case performance.
8.3 Safety¶
Clinical risks, side effects, and effectiveness of risk controls, including evidence that use-related controls work under normal conditions of use.
8.4 Benefit-risk¶
Weigh the demonstrated benefits against residual risks, compared with the state of the art.
8.5 GSPR conformity¶
| GSPR | Evidence | Conclusion |
|---|---|---|
| 1 | ||
| 8 |
8.6 Claims check (Art. 7)¶
Confirm every claim in the intended purpose, IFU, labelling, website, and marketing material is supported. List claims changed or withdrawn.
9. Consistency with Other Documentation¶
- Risk management file (residual risks, benefit-risk conclusion):
- Accompanying information (IFU warnings, limitations, performance information):
- PMS plan and PMCF plan:
- SSCP (class III and implantables) / PSUR:
10. Uncertainties and Post-Market Actions¶
| Uncertainty | Impact | PMS / PMCF action | Timeline |
|---|---|---|---|
11. Conclusions¶
- Safety:
- Performance:
- Clinical benefit:
- Benefit-risk acceptability:
- Sufficiency of clinical evidence for the class and claims:
12. Next Update¶
- Planned date and justification:
- Triggers for an earlier update:
13. References¶
Full citation list, numbered to match references in the text.
14. Signatures¶
- Evaluators (name, role, date):
- QA Lead approval:
- Management Representative approval:
- External evaluator sign-off attached to the approving PR (where applicable):
- Linked approving PR: