Product Classification Decision Template¶
This template shows one workable Markdown-first classification decision record.
Replace the example text below with product-specific content before approval.
Metadata¶
- Decision ID:
- Product ID:
- Jurisdiction:
- Decision Date:
- Owner:
Product Description¶
- Intended Use:
- Target User / Use Environment:
- Claims in Scope:
Classification Analysis¶
- Applicable Rule or Pathway:
- Classification Outcome:
- Rationale:
- Supporting References:
Software Classification Detail (EU MDR / MedDO)¶
Complete this section for medical device software. Guidance: MDCG 2019-11 rev. 1, MDCG 2021-24 rev. 1, IMDRF/SaMD WG/N12 and N81. Base every answer on how the software is realistically used under normal conditions of use (SOP-025 §6.3). Clinician review of the output, disclaimers, and restrictive wording do not by themselves lower the significance of the information.
All Annex VIII rules considered¶
| Rule | Applies? | Resulting class | Reason |
|---|---|---|---|
| Rule 11 | |||
| Other Annex VIII rules that may apply (e.g. Rules 9, 10, 12 for active devices) | |||
| Implementing rule 3.3 (software driving or influencing a device takes that device's class) |
Highest resulting class:
Rule 11, first paragraph: information used for diagnostic or therapeutic decisions¶
- Significance of the information to the healthcare decision (treat or diagnose / drives clinical management / informs clinical management), with reasoning:
- How much users rely on the output in practice; whether there is a realistic opportunity to check the output independently before acting:
- State of the healthcare situation or condition (critical / serious / non-serious), judged by what happens to the patient after an incorrect or late decision rather than the patient's current presentation:
- Worst credible harm from an incorrect, delayed, or unavailable output:
Reference matrix (MDCG 2019-11 rev. 1, adapted from IMDRF/SaMD WG/N12; illustrative, not a substitute for the rule text):
| State of the situation or condition | Treat or diagnose | Drives clinical management | Informs clinical management |
|---|---|---|---|
| Critical | Class III | Class IIb | Class IIa |
| Serious | Class IIb | Class IIa | Class IIa |
| Non-serious | Class IIa | Class IIa | Class IIa |
Rule 11, second paragraph: monitoring of physiological processes¶
- Physiological process monitored:
- Whether variations in vital parameters could result in immediate danger to the patient in the intended clinical context (consider the setting and patient condition, not only the parameter):
- Whether the software also triggers or prioritises clinical action (if so, reassess under the first paragraph):
Rule 11, third paragraph: all other software¶
Software not caught by the first or second paragraph, nor by another rule, is class I. The reference matrix above illustrates only the first paragraph, so class I does not appear in it. - Reason neither the first nor the second paragraph applies:
Follow-Up¶
- Dossier Updates Required:
- Submission or Registration Impact:
- Linked Risk / V&V / Release References:
- Clinical evidence level implications for the clinical evaluation plan (
records/clinical/clinical_evaluation_plan.md):