Equivalence Assessment Template

Implements SOP-025 §6.5(5), MDR Annex XIV Part A section 3, and MDCG 2020-5. Use one record per claimed equivalent device. Equivalence is decided on clinical relevance, not on matching specification sheets.

1. Identification

  • Device under evaluation (name, version, UDI-DI where available):
  • Equivalent device (name, version, manufacturer, CE certificate, UDI-DI):
  • Own earlier version or another manufacturer's device:

2. Access to Data

  • Basis for sufficient access to the equivalent device's technical documentation and clinical data:
  • For class III or implantable devices of another manufacturer, used in order not to perform a clinical investigation (MDR Art. 61(5)):
  • Contract reference allowing ongoing full access to the technical documentation:
  • Evidence that the equivalent device's original clinical evaluation was performed in compliance with the MDR:
  • How both are presented to the notified body:

3. Characteristics Comparison

For each row, state the differences and give the scientific evidence that each difference does not change clinical safety or performance. A difference without evidence fails the row.

Characteristic Device under evaluation Equivalent device Differences Evidence of no clinically significant impact
Technical: design, specifications, principles of operation
Software architecture, algorithm, model, and training data
Inputs (data types, sources, acquisition devices) and outputs
User interface and information presentation
Interoperability and operating platforms
Risk controls and safety features
Biological (usually not applicable to standalone software; state why)
Clinical: condition, indication, severity, stage
Patient population
Intended users and use environment
Clinical workflow and the decision the output supports
Clinical performance and expected benefit

4. Conclusion

  • Equivalence demonstrated: yes / no / partially (state for which claims):
  • Clinical data from the equivalent device to be used, and for which claims:
  • PMCF activities needed to generate data on the device under evaluation: