Equivalence Assessment Template¶
Implements SOP-025 §6.5(5), MDR Annex XIV Part A section 3, and MDCG 2020-5.
Use one record per claimed equivalent device. Equivalence is decided on
clinical relevance, not on matching specification sheets.
1. Identification¶
- Device under evaluation (name, version, UDI-DI where available):
- Equivalent device (name, version, manufacturer, CE certificate, UDI-DI):
- Own earlier version or another manufacturer's device:
2. Access to Data¶
- Basis for sufficient access to the equivalent device's technical documentation and clinical data:
- For class III or implantable devices of another manufacturer, used in order not to perform a clinical investigation (MDR Art. 61(5)):
- Contract reference allowing ongoing full access to the technical documentation:
- Evidence that the equivalent device's original clinical evaluation was performed in compliance with the MDR:
- How both are presented to the notified body:
3. Characteristics Comparison¶
For each row, state the differences and give the scientific evidence that each difference does not change clinical safety or performance. A difference without evidence fails the row.
| Characteristic | Device under evaluation | Equivalent device | Differences | Evidence of no clinically significant impact |
|---|---|---|---|---|
| Technical: design, specifications, principles of operation | ||||
| Software architecture, algorithm, model, and training data | ||||
| Inputs (data types, sources, acquisition devices) and outputs | ||||
| User interface and information presentation | ||||
| Interoperability and operating platforms | ||||
| Risk controls and safety features | ||||
| Biological (usually not applicable to standalone software; state why) | ||||
| Clinical: condition, indication, severity, stage | ||||
| Patient population | ||||
| Intended users and use environment | ||||
| Clinical workflow and the decision the output supports | ||||
| Clinical performance and expected benefit |
4. Conclusion¶
- Equivalence demonstrated: yes / no / partially (state for which claims):
- Clinical data from the equivalent device to be used, and for which claims:
- PMCF activities needed to generate data on the device under evaluation: