Article 61(10) Justification Template¶
Implements SOP-025 §6.9. MDR Art. 61(10) is an alternative way to show
conformity, not an exemption from clinical evaluation. The clinical evaluation,
CER, PMS, and benefit-risk conclusions are still required. "We have not
collected clinical data" is not a justification.
Class III and implantable devices: Art. 61(10) applies without prejudice to
Art. 61(4). It does not remove the requirement for clinical investigations.
Use this template for such a device only together with a recorded Art. 61(4)
to (6) exemption (SOP-025 §6.8(5), §6.9(5)).
- Device class:
- Class III or implantable: yes / no. If yes, exemption relied on and reference:
1. Why Clinical Data Are Not Appropriate¶
Base the argument on the elements named in Art. 61(10):
| Element | Assessment for this device |
|---|---|
| Results of risk management (nature and severity of clinical risks) | |
| Interaction between the device and the patient, users, and clinical process | |
| Intended clinical performance | |
| Claims made by the manufacturer |
Also address: - Are comparable devices in the state of the art normally supported by clinical data? - Would a clinical study add insight into safety or performance that non-clinical methods cannot give? Why not? - How directly does the output determine diagnosis, treatment, or patient management? (The more direct, the harder the justification.)
2. Suitability of the Selected Evidence Sources¶
| Evidence source (e.g. analytical study, MRMC or simulated-use study, usability evaluation, standards conformity, bench testing) | What it can show | What it cannot show |
|---|---|---|
3. How the Evidence Supports Each Clinical Objective¶
| Safety / performance objective (from CEP) | Evidence | Gap remaining |
|---|---|---|
If the device supports clinical decisions, explain how the evidence covers the step from analytical performance to clinical performance: how users read and act on the output, how it fits the clinical workflow, how results differ from one user to another, and use error.
4. Why the Totality of Evidence Is Sufficient¶
Argue sufficiency against GSPR 1, GSPR 8, and the other GSPRs listed in the CEP, compared with state-of-the-art expectations.
5. Post-Market Confirmation¶
- Residual uncertainties:
- PMS / PMCF activities that collect clinical data once the device is in use:
6. Approval Record¶
- Management Representative approval date:
- Linked approving PR: