PMCF Plan Template

Implements MDR Annex XIV Part B section 6 and SOP-025 §6.12. Part of the PMS plan under SOP-014 (MDR Annex III Section 1(b)). Structured on the topics of the MDCG 2020-7 template.

1. Device and Scope

  • Device, versions, and variants covered:
  • Intended purpose and claims being confirmed:
  • Markets and expected user base:

2. Objectives

List the residual uncertainties from the CER this plan addresses, and the standing aims from Annex XIV Section 6.1(a) to (e):

ID Objective Linked CER uncertainty / risk / claim
O-01 Confirm safety and performance over the expected lifetime
O-02 Identify previously unknown side effects and monitor known ones
O-03 Identify emerging risks from evidence
O-04 Keep the benefit-risk ratio acceptable
O-05 Identify systematic misuse or off-label use

3. General PMCF Methods

For example: screening of literature and other sources; analysis of complaints, vigilance data, and user feedback; real-world performance data collected by the software, where lawful under SOP-023.

Method Objective(s) Why this method suits the objective (Annex XIV 6.2(c)) Data source Frequency Responsible

4. Specific PMCF Methods

For example: PMCF study (Art. 74 where it is an investigation), registry, survey, real-world performance study, MRMC re-evaluation. State the protocol reference for each.

Method Objective(s) Why this method suits the objective (Annex XIV 6.2(c)) Design, endpoints, sample size Protocol reference Timeline

A PMCF investigation that adds invasive or burdensome procedures to normal use must be notified to the Member States concerned at least 30 days before it starts (MDR Art. 74(1); SOP-025 §6.8(6)). Include that lead time in the timeline.

Link each specific method to the CER uncertainty it answers. If the plan has no specific method, state which uncertainties general methods resolve and why that is enough (SOP-025 §6.12(2)). A decision that no PMCF is needed at all does not use this template; record it in the PMS plan (SOP-025 §6.12(6)).

5. AI-Specific Monitoring (where applicable)

  • Performance metrics monitored in use and their thresholds:
  • Detection of input-data drift and population shift:
  • Subgroup performance and bias monitoring:
  • Human oversight effectiveness (override rates, disagreement patterns):
  • Link to AI Act post-market monitoring under SOP-024, if in scope:

6. Equivalent and Similar Devices

Evaluation of clinical data on equivalent or similar devices to be monitored:

7. Standards and Guidance

Common specifications, harmonised standards, and guidance applied:

8. Timelines and Reporting

  • PMCF evaluation report frequency:
  • Feed into CER, risk management file, SSCP and PSUR (where applicable):

9. Approval Record

  • Approval date:
  • Linked approving PR: