SOP 025 ClinicalEvaluation

1. Purpose

Define how the organisation plans, conducts, documents, and maintains the clinical evaluation of each medical device software product, and how post-market clinical follow-up (PMCF) keeps that evaluation current. The procedure implements Regulation (EU) 2017/745 (MDR) Article 61 and Annex XIV. For Switzerland, the Medical Devices Ordinance (MedDO, SR 812.213) requires the same clinical evaluation: Art. 21(4) makes it part of demonstrating conformity with the general safety and performance requirements, and Art. 46(3) requires manufacturers to conduct it in accordance with MDR Art. 61, including amendments adopted by delegated acts, and Annex XIV, and to update it with PMCF results. It also supplies the clinical evaluation part of design validation required by ISO 13485:2016 clause 7.3.7.

Clinical evaluation is a continuous, evidence-based argument. It is not a document assembled once before submission. Its output must let a qualified reviewer who did not take part in it follow how each claim about safety, performance, and clinical benefit is supported. Where the evidence does not support a claim, the claim must change.

2. Scope

Applies to every product that qualifies as a medical device (including software as a medical device and software driving or influencing a device) in a jurisdiction recorded in matrices/regulatory_market_scope.yml, for every class, from the start of design and development until the product is retired.

The steps apply to every class, but their depth does not. Scale the volume and type of evidence to the class, the novelty of the device, the severity of the clinical risks, and how directly the output affects patient management, and record that reasoning in the CEP. A class I device with a narrow, well-established purpose may need little more than a focused literature and state of the art review; a novel class IIb or III device usually needs data generated for that device.

Software components without their own medical intended purpose (libraries, APIs, algorithm modules) are evaluated as part of the device they are integrated into. A component placed on the market on its own, with its own intended purpose, is evaluated as a device for each claimed purpose and each combination with other devices for which compatibility is claimed.

In-house devices under MDR Art. 5(5) follow records/mdf/in_house_health_institution_justification_template.md; this SOP applies to them only where the health institution chooses to use it to show conformity with the applicable Annex I requirements. Performance evaluation of in vitro diagnostic software under Regulation (EU) 2017/746 is out of scope.

3. Inputs

  • Approved intended use and claims (records/mdf/intended_use_template.md)
  • Classification decision (records/mdf/classification_decision_template.md)
  • Device description, including variants, configurations, and lifetime
  • Risk management file under SOP-018, including benefit-risk inputs
  • Usability engineering file under SOP-019
  • Verification and validation evidence under SOP-008 and WI-001
  • Post-market surveillance data under SOP-014, complaints under SOP-012, and incidents under SOP-013
  • For AI-enabled software: data governance and model documentation under SOP-022 and, where in scope, SOP-024

4. Outputs

  • Clinical evaluation plan (CEP), including the clinical development plan
  • State of the art (SotA) assessment
  • Literature search protocol and report
  • Appraisal of every data set considered
  • Equivalence assessment or Art. 61(10) justification, where used
  • Clinical evaluation report (CER)
  • PMCF plan and PMCF evaluation report, or a justified decision that PMCF activities beyond PMS are not needed
  • Updates to the risk management file, accompanying information, and claims

5. Roles and Responsibilities

Role Responsibilities
QA Lead Owns this SOP and each product's clinical evaluation file; confirms evaluator competence; checks that the CER, risk file, PMS/PMCF documents, and accompanying information stay consistent; approves every clinical evaluation record.
Management Representative Co-approves with the QA Lead the CEP, CER, PMCF plan, any equivalence assessment, and any use of Art. 61(10); confirms evaluator competence where the QA Lead is also an evaluator; provides resources; receives CER conclusions in management review.
Engineering Lead Supplies device description, technical and analytical performance evidence, change impact information, and data for AI-enabled functions.
Usability Lead Supplies evidence on use-related risk controls and on user performance with the device under representative conditions.
Clinical evaluator(s) Plan and perform the evaluation and author the CER. Not a QMS role: a competence requirement assigned per product under §6.1 and recorded in the CEP. May be employees or external experts.

6. Procedure

6.1 Evaluator Competence and Independence

  1. For each product, name the people who will plan, perform, and review the clinical evaluation. Together they must cover: the clinical field and current clinical practice for the condition concerned; research methodology, including study design and biostatistics; information management for systematic literature searches; regulatory requirements; and the device technology, including AI/ML where present.
  2. Record each evaluator's qualifications and experience and a signed declaration of interests using records/clinical/clinical_evaluator_declaration_template.md. A declared interest is not automatically disqualifying; record how it is managed.
  3. The QA Lead confirms competence before the CEP is approved. Nobody confirms their own competence: where the QA Lead is also an evaluator, the Management Representative confirms instead. Guidance on expected qualifications: MEDDEV 2.7/1 rev. 4 section 6.4.
  4. External evaluators who are not in matrices/signer_registry.json sign their declaration, and their sign-off of the CEP or CER, outside GitHub. Attach the signed documents and a dated CV to the record's approving PR, and record there which internal signer checked them.

6.2 Define the Device and Its Clinical Context

  1. The intended purpose must be stable and specific before conformity assessment starts. It states the medical purpose, indications and contraindications, patient population, intended users and their expertise, use environment, software inputs and outputs, and how the output is used in the clinical workflow (Art. 2(12); Annex II 1.1). Use records/mdf/intended_use_template.md.
  2. Describe normal conditions of use as they will realistically occur: user experience levels, workload, workflow variants, input variability, and reasonably foreseeable misuse. An intended purpose that only holds if users behave perfectly, or if supervision is stricter than sites actually provide, is not acceptable.
  3. Describe where the software sits in the patient pathway, what it changes in that pathway, and how that change is expected to lead to clinical benefit. A workflow diagram is recommended.
  4. Identify novelty: new technology, new clinical application, new population, changed workflow, or new user interaction. Each novel aspect must be addressed explicitly in the evaluation strategy.
  5. State the expected lifetime and the factors that may change safety or performance over it: software updates, platform and dependency changes, changes in clinical practice, drift in input data, and adaptive behaviour.
  6. A significant change to the intended purpose is a design change under SOP-009. It requires reassessment of classification, risk, and the clinical evaluation, not an editorial update.

6.3 Classification Check

  1. Confirm the classification decision covers every applicable MDR Annex VIII rule, not only Rule 11, and applies the highest resulting class.
  2. For Rule 11, assess the significance of the information to the healthcare decision and the seriousness of the healthcare situation or condition, using MDCG 2019-11 rev. 1 and IMDRF/SaMD WG/N12. Base the assessment on realistic use: a clinician making the final decision, disclaimers, or restrictive wording do not by themselves lower the significance of the information.
  3. Record the rationale in records/mdf/classification_decision_template.md. The class sets the level of clinical evidence expected.

6.4 Clinical Evaluation Plan (Annex XIV Part A 1(a))

Prepare and approve a CEP using records/clinical/clinical_evaluation_plan_template.md before evidence is identified. The CEP shall contain, as a minimum: 1. The general safety and performance requirements (GSPRs) that need clinical data. GSPR 1 and GSPR 8 always apply; assess others such as risk control effectiveness, usability, and performance over lifetime. 2. The intended purpose, target groups, indications, and contraindications. 3. The intended clinical benefits, each with specified, measurable clinical outcome parameters. 4. The methods for examining the qualitative and quantitative aspects of clinical safety, with explicit reference to residual risks and side effects. 5. The parameters and acceptance criteria used to judge benefit-risk acceptability, derived from the state of the art. 6. The evidence strategy: the evidence chain in §6.6, the data sources to be used, whether equivalence or Art. 61(10) is intended, and the appraisal method. 7. A clinical development plan describing how evidence will be generated before and after market placement, with milestones and acceptance criteria, and a link to the PMCF plan. 8. How benefit-risk questions raised by particular components (medicinal substances, non-viable tissues of human or animal origin) are addressed. For standalone software this is normally "not applicable"; say so and give the reason, because the item is checked.

Approve the CEP in two steps so that acceptance criteria are never set after seeing the device's own results. The first approval covers scope, the questions to answer, and the state of the art search protocol. Once the state of the art assessment (§6.5(4)) is done, a CEP revision adds the benchmarks and acceptance criteria, and must be approved before any device-specific data set is appraised under §6.7. Any later change to an acceptance criterion needs a recorded reason in the CEP revision history and is reported in the CER.

6.5 Identify Data and Establish the State of the Art (Annex XIV Part A 1(b))

  1. Identify all available data relevant to the device, its equivalent device where used, and the state of the art: clinical investigations, scientific literature, manufacturer-held pre-clinical and clinical data, PMS and PMCF data, registries, usability and simulated-use studies, and retrospective performance studies.
  2. Run literature searches against a written protocol (records/clinical/literature_search_protocol_template.md) with defined questions, databases, search strings, inclusion and exclusion criteria, and date ranges, so that a second person could repeat the search. Separate searches are normally needed for the state of the art, for the device under evaluation (and any equivalent device), and for specific risks, features, or populations.
  3. Include favourable and unfavourable data. Excluding data requires a documented reason.
  4. Write the state of the art assessment using records/clinical/state_of_the_art_assessment_template.md. It describes the condition and current clinical management, alternative technologies and comparable devices, their claims and performance, the risks of the existing care pathway, and from these derives the benchmarks and acceptance criteria used in §6.4(5). Record where each benchmark came from and why it is an appropriate expectation. The state of the art is generally acknowledged good practice, not necessarily the newest technology (MDCG 2020-1).
  5. Equivalence may only be claimed when technical, biological, and clinical characteristics are shown to be equivalent and the manufacturer has sufficient access to the other device's data (Annex XIV Part A 3; Art. 61(5) for class III and implantable devices; MDCG 2020-5). For software, technical and clinical equivalence usually needs detailed justification, including algorithms, inputs, outputs, user interface, and workflow. Use records/clinical/equivalence_assessment_template.md. Any difference must be shown, with evidence, not to change clinical safety or performance. For a class III or implantable device that relies on equivalence to another manufacturer's device in order not to perform a clinical investigation, Art. 61(5) adds two conditions, both to be evidenced to the notified body: a contract giving ongoing full access to that device's technical documentation, and an original clinical evaluation performed in compliance with the MDR.

6.6 Evidence Chain for Software

Assess and report each link separately. Evidence for one link is not evidence for the next (MDCG 2020-1; IMDRF/SaMD WG/N41). 1. Valid clinical association: the output is associated with the targeted clinical condition or physiological state. 2. Technical (analytical) performance: the software reliably and accurately produces the intended output from the input data. Choose metrics relevant to the clinical task; justify why a chosen metric reflects clinically meaningful performance. 3. Clinical performance: using the device, in its intended workflow and with its intended users, achieves the intended effect on the clinical process. 4. Clinical benefit: the positive impact on patient health, patient management, or public health (Art. 2(53)).

Analytical testing on previously collected clinical datasets produces technical or analytical data about the algorithm. It does not, by itself, produce clinical data within Art. 2(48), which comes from the clinical use of the device. Label every data set accordingly in the appraisal.

6.7 Appraise the Data (Annex XIV Part A 1(c))

  1. Define the appraisal method in the CEP and apply it consistently to every data set using records/clinical/clinical_data_appraisal_template.yml.
  2. For each data set record at least: relevance to the device, intended purpose, users, population, and environment; representativeness of normal conditions of use; methodological quality, including study design, sample size, statistics, and bias; the type of data (clinical data, technical or analytical data, or other); which claims, endpoints, or risks it contributes to; and its overall weight.
  3. Use a recognised hierarchy of evidence to weight data, such as MDCG 2020-6 Appendix III together with MEDDEV 2.7/1 rev. 4 section 9 and appendix 6. MDCG 2020-6 was written for legacy devices; its ranking is used here by analogy, and the CEP states that.
  4. Complaint and vigilance data are clinical data under Art. 2(48), but under-reporting makes them a weak source: an absence of complaints does not show that the device is safe or performs as claimed. Use them to detect safety trends and as support only. They are never the sole basis for concluding conformity with the GSPRs (MDCG 2020-6 Appendix III); pair them with appraised clinical data or proactive PMCF data.
  5. Summarise the appraisal as gaps, uncertainties, and open questions for §6.8.

6.8 Generate New Data (Annex XIV Part A 1(d))

  1. Where the appraisal shows that the evidence is not sufficient, plan additional activities. Choose the activity type for the question it has to answer, for example: clinical investigation; prospective or retrospective clinical performance study; multi-reader multi-case (MRMC) or other simulated-use study; summative usability evaluation; registry or real-world data study.
  2. Write a protocol before execution with justified objectives, clinically meaningful endpoints linked to the intended purpose and state of the art, representative conditions of use (or justified limitations), a justified sample size, pre-defined statistical methods and acceptance criteria, unbiased data collection, and follow-up duration.
  3. Clinical investigations follow MDR Articles 62 to 82, Annex XV, and ISO 14155:2020. PMCF investigations also follow Art. 74. Studies with human participants require the applicable ethics and legal approvals.
  4. Report results against the protocol, including deviations, unmet endpoints, limitations, and bias, and their effect on conclusions.
  5. For class III and implantable devices, clinical investigations are required (Art. 61(4)) unless one of the exemptions in Art. 61(4) to (6) applies, such as a modification of the manufacturer's own marketed device whose equivalence the notified body has endorsed, equivalence under the Art. 61(5) conditions, or a device type covered by Art. 61(6). Record in the CEP either the planned investigation or the exemption relied on and the evidence for it (MDCG 2023-7). Where the device is a class III implantable device, or a class IIb active device that administers or removes a medicinal product (Rule 12), plan for the notified body's clinical evaluation consultation procedure (Art. 54).
  6. A PMCF investigation that adds invasive or burdensome procedures to normal use (for example extra imaging, sampling, or visits to establish a reference result) must be notified to the Member States concerned at least 30 days before it starts, with the Annex XV Chapter II documentation (Art. 74(1)). Build that lead time into the PMCF plan.

6.9 Article 61(10) Pathway

  1. Art. 61(10) may be used only where demonstration of conformity based on clinical data is not appropriate. Absence of clinical data, or difficulty collecting it, is not a justification.
  2. Record the justification using records/clinical/article_61_10_justification_template.md. It must show: (a) why clinical data are not appropriate, based on the risk management results, the device's interaction with the patient and the clinical process, the intended clinical performance, and the claims; (b) that the non-clinical evidence sources selected are suitable; (c) how they support each clinically relevant safety and performance objective; and (d) why the totality of evidence is sufficient for the relevant GSPRs.
  3. The more directly the output determines diagnosis or treatment, the harder this justification becomes. For decision-support software, consider whether simulated-use, MRMC, or workflow studies are needed to bridge from analytical performance to clinical performance.
  4. The Management Representative approves any use of Art. 61(10). Clinical data that becomes available after market placement, through PMS or PMCF, must still be collected and evaluated.
  5. Art. 61(10) applies without prejudice to Art. 61(4). For a class III or implantable device it does not remove the requirement for clinical investigations; only an exemption under Art. 61(4) to (6) does (§6.8(5)).

6.10 Analyse and Conclude (Annex XIV Part A 1(e))

  1. Analyse the whole body of appraised evidence against the CEP endpoints, claims, GSPRs, and state-of-the-art benchmarks. Conclusions rest on the totality of evidence, not on one data set.
  2. Confirm every claim made anywhere (intended purpose, labelling, IFU, website, marketing material, technical documentation) is supported by the evidence (Art. 7). Remove or reword unsupported claims.
  3. Evaluate clinical risks together with the effectiveness of risk controls. For controls that depend on how users notice, understand, and act on information (warnings, alerts, confirmations, workflow steps), technical verification that the control exists is not sufficient; use usability, simulated-use, or clinical evidence that it works under normal conditions of use.
  4. Decide whether the benefit-risk ratio is acceptable compared with the state of the art. Record residual uncertainties and assign each one to PMS or PMCF (§6.12).
  5. Keep conclusions consistent with the risk management file under SOP-018. Residual risks and limitations must be reflected in the accompanying information.

6.11 Clinical Evaluation Report (Art. 61(12); Annex XIV Part A 4)

  1. Write the CER using records/clinical/clinical_evaluation_report_template.md. Its structure follows the topics a notified body covers in the clinical evaluation assessment report (MDCG 2020-13), so each topic can be found and verified.
  2. The CER summarises and critically appraises sources rather than reproducing them, and references each source so that a reviewer can sample it. It reports favourable and unfavourable findings.
  3. The CER identifies the software version evaluated and is signed by the evaluators and approved by the QA Lead and the Management Representative.

6.12 Post-Market Clinical Follow-up (Annex XIV Part B)

  1. Prepare a PMCF plan using records/clinical/pmcf_plan_template.md as part of the PMS plan under SOP-014 (Annex III Section 1(b)). The plan targets the residual uncertainties from §6.10, with research questions, general and specific methods, rationale, and timelines (MDCG 2020-7).
  2. Choose methods against the aims in Annex XIV Section 6.1. General methods (user feedback, complaints, literature screening) rarely settle a question the pre-market evidence left open. Plan a specific method, such as a registry or PMCF study (Annex XIV Section 6.2(b)), for each CER uncertainty that general methods cannot answer: little pre-market data for a sub-population or setting, claims that rest mainly on equivalence or on Art. 61(10), or a model whose behaviour on the real input population is not yet known. Where the plan has no specific method, state why in the plan.
  3. Software can often collect real-world performance and use data directly. Where lawful and proportionate under SOP-023, use it to confirm pre-market assumptions.
  4. Report results in a PMCF evaluation report (records/clinical/pmcf_evaluation_report_template.md; MDCG 2020-8). The PMCF evaluation report forms part of the CER and the technical documentation (Annex XIV Section 7). Feed its conclusions into the CER, the risk management file (Annex XIV Section 8) and, where applicable, the summary of safety and clinical performance (SSCP, Art. 32) and periodic safety update report (PSUR, Art. 86).
  5. The PMCF plan also states why each chosen general and specific method is suitable for the question it serves (Annex XIV Section 6.2(c)), not only why other methods were left out.
  6. Concluding that no PMCF is needed at all is a separate decision from having a PMCF plan without specific methods. Record it in the PMS plan with its justification (Annex III Section 1(b)): which residual uncertainties exist, why PMS data alone are enough to address them, and why the device's class, novelty, and evidence base make that acceptable. The QA Lead and Management Representative approve it, and it is re-examined at every CER update.

6.13 Updating the Clinical Evaluation (Art. 61(11))

  1. Update the clinical evaluation and its documentation throughout the lifecycle with clinical data from PMCF and PMS (Art. 61(11)). Set the CER update frequency in the CEP and justify it by class, risk, novelty, and how fast the state of the art changes. For class III and implantable devices, the PMCF evaluation report and, where indicated, the SSCP are updated at least annually. Review the CER and PMCF results at least as often as the PSUR is updated (Art. 86(1): at least annually for class IIb and III, at least every two years for class IIa), so the PSUR never reports clinical conclusions the CER does not contain.
  2. Update it out of cycle when there is a significant safety signal, a PMCF result, a change in the state of the art that affects the benchmarks, a design change that may affect clinical performance or safety under SOP-009, or a change to intended purpose or claims.
  3. The CER always reflects the current released device version. Superseded versions remain retrievable under SOP-001.
  4. Screen every software change raised under SOP-009 for clinical impact, and record the outcome in the change record as one of: (a) no effect on clinical performance, safety, intended purpose, or claims, with the reason (for example a refactoring with unchanged outputs, or a dependency patch with passing regression tests); (b) a CER addendum that records the new version and why existing evidence still covers it; (c) new analytical or clinical performance evidence before release, such as retesting a retrained or re-thresholded model on the locked test set under §6.14(6); (d) a full CER update. Changes to a model, decision threshold, input type, output, or intended user always need at least (b), with the reasoning for not choosing (c).
  5. Where the conformity assessment route requires it, inform the notified body before introducing changes that could affect safety, performance, or the conditions of use (for example MDR Annex IX Section 4.10 or Annex X Section 5.2). The outcome of the screening in item 4 is the input to that decision. Where notified-body approval is required, the changed device must not be released until that approval has been obtained and its evidence linked in the change record.

6.14 AI-Enabled Software

In addition to the steps above: 1. Report error modes over the reasonably foreseeable range of inputs, including atypical and incomplete inputs, as well as summary metrics. An error mode that the user would not notice is a risk to control under GSPR 4 in the order the GSPR sets: design first, then protective measures, and only then information for users. 2. Assess the representativeness of training, tuning, and test data against the intended population and environment, and reflect known data limitations in the intended purpose, warnings, and limitations. 3. For generative models, show that outputs stay consistent and appropriate for the intended purpose across foreseeable inputs. 4. For models that change after market placement, show how performance changes are detected and controlled, and why the lifecycle approach chosen is appropriate compared with controlled versioned updates. 5. Evidence produced under SOP-022 and, where applicable, SOP-024 for data governance, transparency, human oversight, robustness, and post-market monitoring can be reused in the clinical evaluation after appraisal under §6.7. Meeting AI Act obligations is not in itself evidence of MDR conformity, and the reverse also holds (MDCG 2025-6). 6. Performance studies of an AI model must show: (a) how the reference standard (ground truth) was established, by whom, whether they were blinded to the device output, and how disagreement between them was measured and resolved; (b) that no patient, and where relevant no site, appears in both the data used to develop or tune the model and the data used to test it, and that the test set was fixed before the final model was evaluated and not reused to tune it; (c) how performance holds across sites, acquisition equipment and protocols, and patient subgroups, and how the prevalence of the condition in the intended care setting changes predictive values compared with the study population.

7. Required Records

  • Clinical evaluator declarations (records/clinical/clinical_evaluator_declaration_template.md)
  • Clinical evaluation plan (records/clinical/clinical_evaluation_plan_template.md)
  • Literature search protocol and report (records/clinical/literature_search_protocol_template.md)
  • State of the art assessment (records/clinical/state_of_the_art_assessment_template.md)
  • Clinical data appraisal (records/clinical/clinical_data_appraisal_template.yml)
  • Equivalence assessment, where used (records/clinical/equivalence_assessment_template.md)
  • Art. 61(10) justification, where used (records/clinical/article_61_10_justification_template.md)
  • Protocols and reports for any data-generating activity under §6.8
  • Clinical evaluation report (records/clinical/clinical_evaluation_report_template.md)
  • PMCF plan and PMCF evaluation report (records/clinical/pmcf_plan_template.md, records/clinical/pmcf_evaluation_report_template.md)

All records are kept in the product's medical device file under SOP-007.

8. Traceability

Standard Clause Control in this SOP
Regulation (EU) 2017/745 Art. 61(1) GSPR conformity and level of clinical evidence (§6.4, §6.10).
Regulation (EU) 2017/745 Art. 61(1) third subparagraph, Art. 61(3), Annex XIV Part A Section 1 Plan, identify, appraise, generate, analyse (§6.4 to §6.10).
Regulation (EU) 2017/745 Art. 61(4) to (6), Annex XIV Part A 3 Clinical investigations for class III and implantables; equivalence (§6.5(5), §6.8(5), §6.9(5)).
Regulation (EU) 2017/745 Art. 54 Clinical evaluation consultation procedure planned where applicable (§6.8(5)).
Regulation (EU) 2017/745 Art. 74(1) PMCF investigation notification (§6.8(6)).
Regulation (EU) 2017/745 Art. 86(1) CER and PMCF review aligned with PSUR cadence (§6.13(1)).
Regulation (EU) 2017/745 Annex IX 4.10, Annex X 5.2 Clinical impact screening of changes; notified-body notification and pre-release approval where required (§6.13(4), (5)).
Regulation (EU) 2017/745 Art. 61(10) Justification for non-clinical evidence (§6.9).
Regulation (EU) 2017/745 Art. 61(11), (12); Annex XIV Part A 4 Updates and CER (§6.11, §6.13).
Regulation (EU) 2017/745 Annex XIV Part B; Annex III Section 1(b) PMCF, including method rationale and any decision that PMCF is not needed (§6.12).
MedDO (SR 812.213) Art. 21(4), Art. 46(3) Swiss requirement for clinical evaluation under MDR Art. 61 and Annex XIV, updated with PMCF results (§1, §6.4 to §6.13).
Regulation (EU) 2017/745 Art. 7 Claims supported by evidence (§6.10(2)).
ISO 13485:2016 7.3.7 Clinical evaluation as part of design and development validation.
ISO 14971:2019 7.4, 8 Benefit-risk inputs aligned with the risk management file (§6.10(5)).

The clause-level mapping is maintained in matrices/mdr_clinical_evaluation_gap_analysis.yml.

  • QM-001 Quality Manual
  • SOP-001 Document and Record Control
  • SOP-004 Management Review
  • SOP-007 Medical Device File Control
  • SOP-008 Design and Development Control
  • SOP-009 Change Management
  • SOP-011 Competence, Training, and Awareness
  • SOP-012 Feedback and Complaint Handling
  • SOP-013 Regulatory Incident Reporting
  • SOP-014 Post-Market Surveillance
  • SOP-018 Risk Management (ISO 14971:2019)
  • SOP-019 Usability Engineering (IEC 62366-1:2015+A1:2020)
  • SOP-020 Software Lifecycle, Configuration, and Release Management (IEC 62304:2006+A1:2015)
  • SOP-022 AI Management System (ISO/IEC 42001:2023)
  • SOP-023 Data Protection and Privacy Management
  • SOP-024 AI Act Provider QMS (EN 18286:2026)
  • WI-001 Verification and Validation Execution
  • records/clinical/ record family
  • matrices/mdr_clinical_evaluation_gap_analysis.yml

10. Revision History

Revision Revision Date Change Summary
R00 2026-11-19 Initial release of the clinical evaluation and PMCF procedure under MDR Art. 61 and Annex XIV, with software-specific controls for intended purpose, classification, the evidence chain from valid clinical association to clinical benefit, the Art. 61(10) pathway, AI-enabled software, Art. 61(4) to (6) investigation and exemption decisions, PMCF notification and PSUR cadence, clinical impact screening of software changes, and two-step CEP approval, with MedDO Art. 21(4) and Art. 46(3) cited for Switzerland.